tACS - transcranial Alternating Current Stimulation

Transcranial Alternating Current Stimulation generates an alternating current for neuro stimulation.

These are the settings you can alter in tACS mode.

tACS - mode settings

Current 0.1 to 2.0mA This is the maximum current used in the session
Duration 1-40 minutes The duration of the neuro stimulation session.
Bipolar OFF / ON Electrodes change from anode to cathode
Max Voltage 10,15,20,25,30,35,40,45,50,55,60V Limit the maximum voltage - can prevent target current being reached if impedance of electrodes and wave path too high.
Sham OFF / BLIND / ON Sham mode current is started but then turned off. Blind sham it can be on or off.
Duration of real current in sham mode 0, 10, 20, 30, 40, 50 seconds The duration that real current is applied before sham mode turns current off
Current Offset 0.0 to 1.7mA The minimum current value
Frequency 0.1, 0.2, 0.3, 0.4, 0.5, 0.6, 0.7, 0.8, 0.9Hz 
1, 2, 3, 4, 5, 6, 7, 8, 9 Hz 
10, 20, 30, 40, 50, 60, 70, 80, 90 Hz 
100, 150, 200, 250, 300 Hz
The wave frequency in cycles per second
     

This is how to change the tACS settings on your foc.us v2

Select tACS mode type

Set sham OFF

Still need help? Contact Us Contact Us